Original ArticleIndian Journal of Pharmaceutical Education and ResearchVol. 59 | Issue 2s | 2025 | pp. s689–s704Open access
Quality by Design (QbD) Based Development and Validation of RP-HPLC Method for Simultaneous Estimation of Metformin and Teneligliptin in Bulk and their Pharmaceutical Formulation
- 1*,
- 2*,
- 2*,
- 2*,
- 2*,
- 2*
- 1 Department of Pharmaceutical Chemistry and Quality Assurance, L. M. College of Pharmacy, Ahmedabad, Gujarat, INDIA.
- 2 Faculty of Pharmacy, Dharmsinh Desai University, Nadiad, Gujarat, INDIA.
Published in Indian Journal of Pharmaceutical Education and Research
Correspondence: Jagatkumar Upadhyay
Department of Pharmaceutical Chemistry and Quality Assurance, L. M. College of Pharmacy, Ahmedabad, Gujarat, INDIA.
Email: jagat.upadhyay@lmcp.ac.in
Copyright: © 2025 Manuscript Technomedia. This is an open access article.
- Published:
- Apr 23, 2025
- Received:
- Mar 18, 2023
- Accepted:
- Dec 23, 2024
How to cite
Upadhyay, J., Padhiyar, S., Jivani, K., Patel, T., Bhavsar, V., & Prajapati, V. (2025). Quality by Design (QbD) Based Development and Validation of RP-HPLC Method for Simultaneous Estimation of Metformin and Teneligliptin in Bulk and their Pharmaceutical Formulation. Indian Journal of Pharmaceutical Education and Research, 59(2s), s689–s704. https://doi.org/10.5530/ijper.20253717
Abstract
Aim:
To develop and validate simple, accurate, precise, sensitive and robust Reversed Phase High Performance Liquid Chromatographic (RP-HPLC) method for simultaneous estimation of metformin and teneligliptin in bulk and their pharmaceutical formulation by using Quality by Design (QbD) approach.
Materials and Methods:
The factor screening studies were performed using 2 level 4 factor (16 run) full factorial design. C18 Gracesmart column (150×4.6 mm, 5 μ) was saturated with mobile phase acetate buffer pH 4: acetonitrile (56.4: 43.6 v/v). Mobile phase was pumped at flow rate 1 mL/min. Both the drugs metformin and teneligliptin detected by diode array detector at wavelength 250 nm.
Results:
Retention time for metformin and teneligliptin were found to be 1.983 min and 3.263 min respectively. Metformin and teneligliptin show linear response in concentration range 50-300 μg/mL and 2-12 μg/mL respectively. % Recovery for metformin and teneligliptin was found to be 100.15%-101.43% and 98.76%-99.48% respectively. In precision study % relative standard deviation was found less than 2 for metformin and teneligliptin. Limit of Detection (LOD) for metformin and teneligliptin was found to be 1.281 μg/ mL and 0.464 μg/mL respectively. Limit of Quantification (LOQ) for metformin and teneligliptin was found to be 3.881 μg/mL and 1.407 μg/mL respectively.
Conclusion:
The proposed method was found to be accurate, precise, sensitive and robust. As a result, it might be successfully employed for the routine analysis of metformin and teneligliptin in bulk and their pharmaceutical dosage forms.
Keywords
Subject
Article metadata
| Title | Quality by Design (QbD) Based Development and Validation of RP-HPLC Method for Simultaneous Estimation of Metformin and Teneligliptin in Bulk and their Pharmaceutical Formulation |
|---|---|
| Authors | Jagatkumar Upadhyay; Shantilal Padhiyar; Kamlesh Jivani; Tejas Patel; Vashisth Bhavsar; Vipul Prajapati |
| Affiliations | Department of Pharmaceutical Chemistry and Quality Assurance, L. M. College of Pharmacy, Ahmedabad, Gujarat, INDIA.; Faculty of Pharmacy, Dharmsinh Desai University, Nadiad, Gujarat, INDIA. |
| Corresponding author | jagat.upadhyay@lmcp.ac.in |
| Journal | Indian Journal of Pharmaceutical Education and Research |
| Volume / Issue | Vol. 59, Issue 2s (2025) |
Also in this issue
- Nanofibers: Promising Novel Drug Delivery Technology for Cosmeceuticalspp. s390–s398
- Casting a Spotlight on Factorial Design: Exploring the Power of DoE for Experiment Screening and Optimization: A Mini Reviewpp. s399–s409
- A Qualitative Study on Lecture Capturing System in Pharmacy Education-Learners’ Perspectivepp. s410–s413
- Formulation and Evaluation of Simvastatin Loaded Nanostructured Lipid Carrier for Topical Drug Deliverypp. s414–s425
- Optimized Vildagliptin Microsphere Formulation and in vitro Characterization Using 23 Factorial Design Analysispp. s426–s435
Readers Also Viewed
Development and Validation of UV/visible Spectrophotometric Method for Estimation of Piroxicam from Bulk and Formulation
Sandip Mohan Honmane, Kunal Rajaram Yadav, Yuvraj Dilip Dange
Apr 23, 2025
Effects of Artificial Intelligence on Academic Performance of Library and Information Science University Students: A Meta-Analysis (2023-2025)
Kayode Sunday John Dada
Aug 6, 2026
Bridging Innovation and Impact: A Multidisciplinary Approach to Contemporary Research Challenges
Mueen Ahmed KK
Aug 11, 2026